ScienceChemistry

Beer-Lambert Law
calculator.

Turn a spectrophotometer reading into a concentration, or work the law backward.

How it's worked out

A spectrophotometer shines light through your sample and compares what comes out with what went in. The ratio is the transmittance. Harvey's Analytical Chemistry/10%3A_Spectroscopic_Methods/10.02%3A_Spectroscopy_Based_on_Absorption) defines absorbance from it with a base 10 log.

A\displaystyle A=−log10⁡T\displaystyle {}= -\log_{10} T

Absorbance has no units. The reason to use it is that it rises in a straight line with concentration, while transmittance falls along a curve.

A\displaystyle A=εlc\displaystyle {}= \varepsilon \, l \, c

Here ε is the molar absorptivity, a property of the compound at one wavelength, in L/(mol·cm). Some sources call it the molar extinction coefficient. The path l is the distance the light travels through the sample, in cm, and c is the concentration in mol/L. Any one of the terms follows from the other three, and the boxes convert µM and mm first.

Two compounds at once. Harvey notes that absorbances add when the compounds do not interact. Measure at two wavelengths and you get two equations in two unknowns.

A1\displaystyle A_1=l(ε1aca+ε1bcb)\displaystyle {}= l(\varepsilon_{1a} c_a + \varepsilon_{1b} c_b)
A2\displaystyle A_2=l(ε2aca+ε2bcb)\displaystyle {}= l(\varepsilon_{2a} c_a + \varepsilon_{2b} c_b)

Worked examples

Sana reads NADH in an enzyme assay

NADH absorbs at 340 nm, and a Redox Report review gives its coefficient as 6.22 L/mmol/cm, which is 6,220 L/(mol·cm). Sana reads 0.50 in a 1 cm cell, so her sample holds 80.4 µM.

The same review checks the other way: 0.1 mM reads 0.622 through 1 cm. With a 1 mm cell, Sana's sample would read only 0.05, which is one reason path length belongs in the sum.

Open this example in the tool

Theo finds ε from a standard he made

His dye has no published coefficient at his wavelength. A 120 µM standard reads 0.84 in a 1 cm cell, so ε is 7,000 L/(mol·cm).

A sample of the same dye that reads 0.42 is then 60 µM, half the reading and half the concentration. One standard makes a quick estimate, and a series of them shows whether the line stays straight.

Open this example in the tool

Ines separates two dyes in one sample

Dye A absorbs strongly at the first wavelength, with ε = 1,000, and weakly at the second, with ε = 100. Dye B does the reverse, with 200 and 800. Readings of 0.50 and 0.40 give 410 µM of A and 449 µM of B.

The sum relies on the two dyes having different spectra. If they absorbed in the same ratio at both wavelengths, the two equations would say the same thing, and no answer would exist.

Open this example in the tool

Reading your result

Each extra 1 of absorbance cuts the light through to a tenth. Absorbance is not a percent, and 1 does not mean all the light is gone.

Absorbance against light
AThroughAbsorbed
0.179.4%20.6%
0.531.6%68.4%
110%90%
21%99%
30.1%99.9%

Where a reading is most precise. Tom O'Haver of the University of Maryland puts the best precision at an absorbance of about 0.5 to 1.0. Below that, the small drop in light is hard to measure. Above it, very little light reaches the detector.

Where the line starts to bend. O'Haver says deviations normally appear above an absorbance of about 2. Shimadzu gives about 4 for an instrument with 0.001 %T of stray light, which is light that reaches the detector without passing through the sample at the set wavelength. So the top limit depends on your instrument, and its specification sheet outranks the rule of thumb here.

If a reading is too high, dilute. The law is proportional in c, so a reading of 2.4 diluted 4 times should read 0.6. Measure the diluted sample, then multiply the concentration back up by the same factor.

Standards beat a single ε. O'Haver recommends a series of standard solutions. A curve made from them still gives accurate results when the law is not obeyed perfectly, and a sample such as a 25% transmittance reading (A = 0.602, or 40.1 µM with ε = 15,000) can then be read against it.

Common mistakes

  • Reading high and trusting it. Shimadzu says stray light always gives lower than actual absorbance, so a dark sample reads too low and the concentration comes out too small. The line looks fine until you check it with a dilution.
  • Using ε from the wrong wavelength. The coefficient belongs to one compound at one wavelength. Harvey adds that polychromatic light, a band of wavelengths instead of one, always bends the line downward.
  • Forgetting the cell. A plate reader or a micro-volume instrument often has a path well under 1 cm. Enter the true path, as a 1 mm path lowers the reading tenfold.
  • Assuming the law holds at any strength. Harvey calls it a limiting law, valid only for low concentrations. At higher ones the molecules no longer act independently, and a weak acid that changes form can shift the reading too.
  • Reading a cloudy sample. Haze or bubbles that scatter light take light away from the detector, and the instrument counts that as absorbance. Filter the sample or spin it down first.

Questions people ask

Yes. An absorbance of 1 lets 10% of the light through, and 2 lets 1% through, so neither means the light has gone. How high you can trust a reading depends on the stray light in your instrument.